Search MEDLINE/PubMed
Title: Correction for eddy current induced Bo shifts in diffusion-weighted echo-planar imaging. Author: Calamante F, Porter DA, Gadian DG, Connelly A. Journal: Magn Reson Med; 1999 Jan; 41(1):95-102. PubMed ID: 10025616. Abstract: The importance of diffusion-weighted MRI in the assessment of acute stroke is well-recognized, and quantitative maps of the apparent diffusion coefficient (ADC) are now widely used. Echoplanar imaging provides a robust method of acquiring diffusion-weighted images free of motion artifact. However, initial experience with clinical MRI systems indicates that calculation of artifact-free ADC maps from a series of echo-planar diffusion-weighted images is not necessarily straight-forward. One of the problems is that frequency shifts resulting from eddy currents can cause misregistration of base diffusion-weighted images. In this study, an on-line correction method that overcomes this problem is described, and phantom and human images that demonstrate the validity of the technique are presented. The method uses a non-phase-encoded reference scan to correct the phase of each echo in the echo train, and can provide ADC maps that are free of misregistration artifacts, without the need for off-line postprocessing.[Abstract] [Full Text] [Related] [New Search]