These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: High-speed data acquisition system and receiver configurations for time-domain radiofrequency electron paramagnetic resonance spectroscopy and imaging. Author: Subramanian S, Murugesan R, Devasahayam N, Cook JA, Afeworki M, Pohida T, Tschudin RG, Mitchell JB, Krishna MC. Journal: J Magn Reson; 1999 Apr; 137(2):379-88. PubMed ID: 10089172. Abstract: Design strategies, system configuration, and operation of a dual-channel data acquisition system for a radiofrequency (RF) time-domain electron paramagnetic resonance (EPR) spectrometer/imager operating at 300 MHz are described. This system wasconfigured to incorporate high-speed analog-to-digital conversion (ADC) and summation capabilities with both internal and external triggering via GPIB interface. The sampling rate of the ADC is programmable up to a maximum of 1 GS/s when operating in a dual-channel mode or 2 GS/s when the EPR data are collected in a single-channel mode. By using high-speed flash ADCs, a pipelined 8-bit adder, and a 24-bit accumulator, a repetition rate of 230 kHz is realized to sum FIDs of 4096 points. The record length is programmable up to a maximum of 8K points and a large number of FIDs (2(24)) can be summed without overflow before the data can be transferred to a host computer via GPIB interface for further processing. The data acquisition system can operate in a two-channel (quadrature) receiver mode for the conventional mixing to baseband. For detection using the single-channel mode, the resonance signals around the center frequency of 300 MHz were mixed with a synchronized local oscillator of appropriate frequency leading to an intermediate frequency (IF) which is sampled at a rate of 2 GS/s. Comparison of quadrature-mode and an IF-mode operation for EPR detection is presented by studying the FID signal intensity across a bandwidth of 10 MHz and as a function of transmit RF power. Imaging of large-sized phantoms accommodated in appropriately sized resonators indicates that IF-mode operation can be used to obtain distortion-free images in resonators of size 50 mm diameter and 50 mm length.[Abstract] [Full Text] [Related] [New Search]