These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Rapid high-performance liquid chromatographic determination of serum gabapentin.
    Author: Jiang Q, Li S.
    Journal: J Chromatogr B Biomed Sci Appl; 1999 Apr 30; 727(1-2):119-23. PubMed ID: 10360430.
    Abstract:
    Gabapentin (GBP) is a new antiepileptic drug approved for clinical treatment of partial seizures in the USA. Serum GBP concentrations in 283 patients were studied using high-performance liquid chromatography with fluorescence detection. The standard curves were linear over a range of 60 ng to 15 microg/ml. The coefficient of variations were 3.4 to 8.8% and 1.4 to 9.8% for intra- and inter-assay studies, respectively. The lower limit of quantitation was 10 ng/ml. Of the 283 patients studied, 72.5% had GBP levels between 2 and 10 microg/ml, 14.8% were below 2 microg/ml and 12.7% above 10 microg/ml. The mean+/-S.E. of GBP in 283 patients was 5.38+/-0.23 microg/ml. Peak concentrations of more than 15 microg/ml and trough levels as low as 0.1 microg/ml were not uncommon. The method described was rapid, simple, highly sensitive and reproducible. Other antiepileptic drugs and endogenous compounds did not interfere with the assay.
    [Abstract] [Full Text] [Related] [New Search]