These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Direct digital magnification mammography with a large-surface detector made of amorphous silicon]. Author: Hermann KP, Hundertmark C, Funke M, von Brenndorff A, Grabbe E. Journal: Rofo; 1999 May; 170(5):503-6. PubMed ID: 10370416. Abstract: PURPOSE: To investigate the image quality of a new direct digital mammography system using a large-area amorphous silicon X-ray detector in a phantom study. MATERIALS AND METHODS: The contrast-detail resolution as a function of the tube voltages, the magnification factors and the mean glandular doses were investigated using dedicated test objects. RESULTS: The contrast-detail resolution was significantly improved in comparison with conventional screen-film mammography. Usually, the doses necessary to obtain these high-quality survey mammograms were smaller. CONCLUSIONS: By combining the direct magnification technique and a digital flat panel detector the limited spatial resolution of such image receptors can be overcome. With this direct digital mammography technique, a digital image was directly captured without an intermediate step of optical or mechanical scan.[Abstract] [Full Text] [Related] [New Search]