These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Performance of open-fronted microbiological safety cabinets: the value of operator protection tests during routine servicing. Author: Osborne R, Durkin T, Shannon H, Dornan E, Hughes C. Journal: J Appl Microbiol; 1999 Jun; 86(6):962-70. PubMed ID: 10389245. Abstract: The performance of class I and II microbiological safety cabinets over 7 years, employed in a force-ventilated containment level 3 (CL-3) laboratory, is described. Operator Protection (OP) provided by the cabinets, assessed by still and latterly limited 'in-use' KI-Discus tests, showed no overall deterioration during the review period. Comparisons show that a selected class II unit, but not a second, and a new class II MSC in a recently commissioned, similar CL-3 facility, provide the same order of OP as a class I cabinet. From the experiences described, it is strongly recommended that OP tests (OPTs) should be part of the routine servicing regime to ensure that cabinets meet required performance levels, and additionally to allow detection and rectification of poor containment, particularly where induced by environmental factors. The value of OPTs is discussed with reference to certain national standards.[Abstract] [Full Text] [Related] [New Search]