These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Feasibility of simultaneous stress 99mTc-sestamibi/rest 201Tl dual-isotope myocardial perfusion SPECT in the detection of coronary artery disease.
    Author: Nakamura M, Takeda K, Ichihara T, Motomura N, Shimizu H, Saito Y, Nomura Y, Isaka N, Konishi T, Nakano T.
    Journal: J Nucl Med; 1999 Jun; 40(6):895-903. PubMed ID: 10452303.
    Abstract:
    UNLABELLED: This study assesses feasibility and diagnostic accuracy of simultaneous stress 99mTc-sestamibi/rest 201 TI dual-isotope myocardial perfusion SPECT with Moore's correction method, in which contamination originating from lead x-rays produced in a collimator was subtracted in the 201TI windows. METHODS: Eighty-one patients with suspected coronary artery disease received exercise 99mTc-sestamibi injection, followed by rest 201TI injection 50 min later, and dual-isotope SPECT was performed (group 1). These results were compared with coronary angiographic findings. Furthermore, to estimate the accuracy of Moore's correction method, 201TI crosstalk into the 99mTc acquisition window (group 2A, n = 20) and 99mTc crosstalk into the 201TI acquisition windows (group 2B, n = 20) were studied. For group 2A, stress 99mTc-sestamibi SPECT (single 99mTc-sestamibi SPECT) was performed, followed by 201TI injection at rest and dual-isotope SPECT acquisition 50 min later. For group 2B, rest 201TI SPECT (single 201TI SPECT) was performed, followed by 99mTc-sestamibi injection at rest and dual-isotope SPECT acquisition 30 min later. RESULTS: Sensitivity and specificity in group 1 were 83% and 99%, respectively, when > or =75% coronary artery narrowing was considered significant. In groups 2A and 2B, SPECT images were divided into 24 segments, and relative regional uptake in each segment was obtained. In group 2A, relative regional uptake of single 99mTc-sestamibi SPECT correlated well with that of dual-isotope SPECT (r = 0.942). In group 2B, relative regional uptake of single 201TI SPECT correlated well with that of dual-isotope SPECT (r = 0.935). Furthermore, in low 201TI uptake segments with relative regional uptake in both single- and dual-isotope SPECT of < or =70%, the degree of concordance between single- and dual-rest 201TI was considered to be high with Bland-Altman analysis and the kappa statistic. Comparison of perfusion defect type demonstrated that, of 22 stress defects within infarct zones, 95% were irreversible and 5% were reversible. In contrast, of 28 stress defects within stenosed vessel zones in noninfarct zones, 89% were reversible and 11% were irreversible (P < 0.0001 versus infarct zones). CONCLUSION: Simultaneous dual-isotope imaging with Moore's correction method is feasible, with acceptable accuracy for detection of coronary artery disease and a small amount of crosstalk into each window.
    [Abstract] [Full Text] [Related] [New Search]