These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Performance characteristics of IMMULITE TPS: a comparison with TPS IRMA.
    Author: Einarsson R, Lei JD, Ullrich A, Van Dalen A.
    Journal: Anticancer Res; 1999; 19(4A):2743-7. PubMed ID: 10470233.
    Abstract:
    This paper describes an automated immunoassay for the measurement of TPS developed for the IMMULITE. The IMMULITE system is a fully automated continous random access analyzer, which uses enzyme-amplified chemiluminiscent as the detection system. The IMMULITE TPS assay is a sequential, two-site chemiluminiscent enzyme immunometric assay designed for the quantitative measurement of TPS in serum. The IMMULITE TPS assay covers a clinical concentration range of 20-2400 U/L, with a lower detection limit of 6 U/L. A serum method comparison (n = 340) to the TPS IRMA assay demonstrates a correlation coefficient of 0.961 and a slope of 1.04. Samples run on IMMULITE TPS assay exhibit linearity upon dilution where recovered values fall within 15% of the theoretical sample value. The intra-assay precision of the IMMULITE TPS assay ranged from 3.3-4.7%, while the interassay precision ranged from 3.9-6.0%. By using different methods (immunoradiometric assay vs chemiluminiscent enzyme-labeled immunometric assay, one-step procedure vs sequential two-step procedure, manual vs fully automated procedure) the two TPS tumor marker tests used in this study gave quite comparable determinations for the sera from the cancer patients.
    [Abstract] [Full Text] [Related] [New Search]