These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Method of reduction of the number of driving system channels for phased-array transducers using isolation transformers.
    Author: Fjield T, Hynynen K.
    Journal: IEEE Trans Biomed Eng; 2000 Jan; 47(1):139-41. PubMed ID: 10646290.
    Abstract:
    Phased-array technology offers an incredible advantage to therapeutic ultrasound due to the ability to electronically steer foci, create multiple foci, or to create an enlarged focal region by using phase cancellation. However, to take advantage of this flexibility, the phased-arrays generally consist of many elements. Each of these elements requires its own radio-frequency generator with independent amplitude and phase control, resulting in a large, complex, and expensive driving system. A method is presented here where in certain cases the number of amplifier channels can be reduced to a fraction of the number of transducer elements, thereby simplifying the driving system and reducing the overall system complexity and cost, by using isolation transformers to produce 180 degrees phase shifts.
    [Abstract] [Full Text] [Related] [New Search]