  • Title: Kinetic analysis of the mass transport limited interaction between the tyrosine kinase lck SH2 domain and a phosphorylated peptide studied by a new cuvette-based surface plasmon resonance instrument.
    Author: de Mol NJ, Plomp E, Fischer MJ, Ruijtenbeek R.
    Journal: Anal Biochem; 2000 Mar 01; 279(1):61-70. PubMed ID: 10683231.
    Abstract:
    We explored the use of a newly developed cuvette-based surface plasmon resonance (SPR) instrument (IBIS) to study peptide-protein interactions. We studied the interaction between the SH2 domain of lck and a phosphotyrosine peptide EPQY*EEIPIYL which was immobilized on a sensor chip. No indications for mass transport limitation (MTL) were observed when standard kinetic approaches were used. However, addition of competing peptide during dissociation revealed a high extent of rebinding. A dissociation rate constant (k(d)) of 0.6+/-0.1 s(-1) was obtained in the presence of large amounts of peptide. A simple bimolecular binding model, applying second-order kinetics for the cuvette system, could not adequately describe the data. Fits were improved upon including a step in the model which describes diffusion of the SH2 domain from the bulk to the sensor, especially for a surface with high binding capacity. From experiments in glycerol-containing buffers, it appeared that the diffusion rate decreased with higher viscosity. It is demonstrated that MTL during association and dissociation can be described by the same diffusion rate. A binding constant (K(D)) of 5.9+/-0.8 nM was obtained from the SPR equilibrium signals by fitting to a Langmuir binding isotherm, with correction for loss of free analyte due to binding. An association rate constant k(a) of 1.1(+/-0.2)x10(8) M(-1) x s(-1) was obtained from k(d)/K(D). The values for k(a) and k(d) obtained in this way were 2-3 orders larger than that from standard kinetic analysis, ignoring MTL. We conclude that in a cuvette the extent of MTL is comparable to that in a flow system.
    [Abstract] [Full Text] [Related] [New Search]