These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of glomerular filtration rate using a dual-detector gamma camera and the geometric mean of renal activity: correlation with the Tc-99m DTPA plasma clearance method.
    Author: Delpassand ES, Homayoon K, Madden T, Mathai M, Podoloff DA.
    Journal: Clin Nucl Med; 2000 Apr; 25(4):258-62. PubMed ID: 10750963.
    Abstract:
    PURPOSE: In this study, the authors studied the use of a dual-detector gamma camera to measure the glomerular filtration rate (GFR). METHODS: Thirty-three patients with a wide range of renal function participated in this study. The GFR was measured using a dual-detector gamma camera by calculating the geometric mean of activity from each kidney and using an outline background. These results were compared with the GFR estimates obtained from Tc-99m DTPA plasma clearance using a multiple blood sample method. RESULTS: Correlation was excellent between GFR estimated using the dual-detector gamma camera and GFR measured using the plasma clearance of Tc-99m DTPA with multiple blood samples (r = 0.89). The correlation was especially strong in children younger than 13 years (r = 0.94). CONCLUSION: Measuring the GFR using a dual-detector gamma camera and calculating the geometric mean of renal activity yields relatively accurate results.
    [Abstract] [Full Text] [Related] [New Search]