These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Goldmann applanation versus hand-applanation and .schiötz indentation tonometry. Author: Krieglstein GK, Waller WK. Journal: Albrecht Von Graefes Arch Klin Exp Ophthalmol; 1975; 194(1):11-6. PubMed ID: 1079117. Abstract: For evaluation of hand-applanation tonometry comparative measurements of intraocular pressure have been carried out using the Goldmann applanation tonometer, the Draeger and Perkins hand-applanation tonometers and the Schiötz indentation tonometer. A series of 52 eyes has been tested. The statistical analysis was based on the Goldmann tonometry as a reference system. The hand-applanation tonometers were found to be in satisfactory agreement with the Goldmann tonometer, whereas the Schiötz tonometer gave significantly low values in the pressure range considered.[Abstract] [Full Text] [Related] [New Search]