These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Tooth enamel EPR dosimetry: sources of errors and their correction. Author: Ivannikov AI, Skvortsov VG, Stepanenko VF, Tsyb AF, Khamidova LG, Tikunov DD. Journal: Appl Radiat Isot; 2000 May; 52(5):1291-6. PubMed ID: 10836445. Abstract: Some of the most important sources of systematic errors in dose determination using tooth enamel EPR spectroscopy and ways of reducing those errors are discussed. Enamel from the outside of the front teeth should not be used for dose determination because of induction of paramagnetic centers by solar light. The accuracy of the method in the low dose range is limited by variation in the shape of the EPR signal of unirradiated enamel, which can be described by an initial intrinsic signal and which varies for different samples with standard deviation of 20-30 mGy. The energy dependence of enamel sensitivity should be taken into account in the form of a correction factor. The value of this factor is estimated at 1.1-1.3 for real radiation fields in radiation contaminated territories. Variation in enamel sensitivity for different samples is shown to be within limits of 10-15% of the average value.[Abstract] [Full Text] [Related] [New Search]