  • Title: In-phantom response of LiF TLD-100 for dosimetry of 192Ir HDR source.
    Author: Pradhan AS, Quast U.
    Journal: Med Phys; 2000 May; 27(5):1025-9. PubMed ID: 10841406.
    Abstract:
    An experiment was carried out to reevaluate the response of LiF TLD-100 rods (1 mmx1 mmx6 mm) at different depths in a water substituting phantom to provide an answer to a prevailing controversy about the over-response of LiF to the softened photon spectra of 192Ir HDR source at depths in phantom due to its photon energy dependence. Claims of some authors that LiF TLDs over-responds by 8.5% at 10 cm depth in phantom, necessitating depth-dependent correction factors even for an 192Ir source and of some others for no over-response were evaluated. The over-response of LiF TLD-100 rods, against a calibrated ion chamber having a photon energy-independent response within 2%, was found to be not exceeding 2.5% at a depth of 10 cm in the phantom as compared to a depth at 1 cm, for a precision of the order of +/- 1% (1sigma) in the TLD measurements. By using ISO equivalent photon beams, photon energy dependence of the dosimeters was evaluated and for LiF TLD-100 rods it was found to be in close agreement (within 3%) with the ratios of mass energy absorption coefficients of LiF and water in the range of effective photon energy from 26 keV to 1.25 MeV. Parameters that could contribute to the discrepancy in the reported values of experimental results have been discussed.