Title: Fixation of periprosthetic femoral shaft fractures occurring at the tip of the stem: a biomechanical study of 5 techniques. Author: Dennis MG, Simon JA, Kummer FJ, Koval KJ, DiCesare PE. Journal: J Arthroplasty; 2000 Jun; 15(4):523-8. PubMed ID: 10884215. Abstract: This study evaluated 5 currently used periprosthetic femoral shaft fracture fixation techniques to determine which technique provided the greatest fixation stability. Periprosthetic fractures in 30 synthetic femurs were fixed with a plate with cables, plate with proximal cables and distal bicortical screws (Ogden concept), plate with proximal unicortical screws and distal bicortical screws, plate with proximal unicortical screws and cables and distal bicortical screws, or 2 allograft cortical strut grafts with cables. These specimens were then tested in 3 physiologic loading modes. The plate constructs with proximal unicortical screws and distal bicortical screws or with proximal unicortical screws, proximal cables, and distal bicortical screws were significantly more stable in axial compression, lateral bending, and torsional loading than the other fixation constructs studied.[Abstract] [Full Text] [Related] [New Search]