These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The precision of a special purpose analog computer in clinical cardiac output determination. Author: Sullivan FJ, Mroz EA, Miller RE. Journal: Ann Surg; 1975 Feb; 181(2):232-8. PubMed ID: 1089394. Abstract: Three hundred dye-dilution curves taken during our first year of clinical experience with the Waters CO-4 cardiac output computer were analyzed to estimate the errors involved in its use. Provided that calibration is accurate and 5.0 mg of dye are injected for each curve, then the percentage standard deviation of measurement using this computer is about 8.7%. Included in this are the errors inherent in the computer, errors due to baseline drift, errors in the injection of dye and acutal variation of cardiac output over a series of successive determinations. The size of this error is comparable to that involved in manual calculation. The mean value of five successive curves will be within 10% of the real value in 99 cases out of 100. Advances in methodology and equipment are discussed which make calibration simpler and more accurate, and which should also improve the quality of computer determination. A list of suggestions is given to minimize the errors involved in the clinical use of this equipment.[Abstract] [Full Text] [Related] [New Search]