These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of vitamin K1 in emulsified nutritional supplements by solid-phase extraction and high-performance liquid chromatography with postcolumn reduction on a platinum catalyst and fluorescence detection. Author: Iwase H. Journal: J Chromatogr A; 2000 Jun 09; 881(1-2):261-6. PubMed ID: 10905710. Abstract: Determination of small amounts of vitamin K1 (0.8 microg/g) in nutritional supplements with high fat content (20 mg/g) was performed by solid-phase extraction and high-performance liquid chromatography (HPLC) with fluorescence detection after reduction on a platinum oxide catalyst. The concentration ratio of plant oils to vitamin K1 (0.8 microg/g) was about 25,000:1. A sample solution was applied to a solid-phase extraction cartridge and vitamin K1 was eluted with ethanol, followed by HPLC. The proposed method was simple, rapid (analysis time: ca. 12 min), sensitive [detection limit: ca. 0.1 pg per injection (100 microl) at a signal-to-noise ratio of 3:1], highly selective and reproducible [relative standard deviation: ca. 1.3%. (n=5)]. The calibration graph of vitamin K1 was linear in the range of 0-2 pg per injection (100 microl). Recovery of vitamin K1 was over 90% by the standard addition method.[Abstract] [Full Text] [Related] [New Search]