These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates. Author: Girouard J, Forest JC, Massé J, Leroux M, Bradburn NC, Noblet TC, Joynes JO, Baum J. Journal: Diabetes Care; 2000 Aug; 23(8):1149-53. PubMed ID: 10937513. Abstract: OBJECTIVE: To evaluate the clinical performance of the Glucometer Elite XL Diabetes Care System in neonatal settings using a multicenter study RESEARCH DESIGN AND METHODS: A total of 388 blood specimens from 333 neonates were included in the study. A capillary or arterial sample was analyzed for determination of glucose with the Glucometer Elite XL system by an attending trained nurse. Through the same sampling site, a specimen was collected and sent to the laboratory for measurement of plasma glucose, bilirubin, and hematocrit. RESULTS: The regression analysis between the results of the Glucometer Elite XL system and comparative methods resulted in the following: Glucometer Elite XL meter = 1.01 x laboratory method + 0.02 mmol/l (n = 388). For the 1.1-4.0 mmol/l plasma glucose range, the regression was Glucometer Elite XL meter = 1.07 x laboratory method + 0.12 mmol/l (n = 150). A difference plot indicated a mean bias of 0.04 mmol/l (95% CI -0.01 to 0.10). No relationship was found between meter glucose biases and hematocrit levels (r = 0.10, P = 0.14). Although a statistically significant correlation existed between bilirubin levels and the glucose meter biases (r = 0.14, P = 0.005), the predicted mean biases were of little clinical significance. CONCLUSIONS: The Glucometer Elite XL system showed a good performance when used in neonatal settings.[Abstract] [Full Text] [Related] [New Search]