  • Title: Use of a microplate scintillation counter as a radioactivity detector for miniaturized separation techniques in drug metabolism.
    Author: Boernsen KO, Floeckher JM, Bruin GJ.
    Journal: Anal Chem; 2000 Aug 15; 72(16):3956-9. PubMed ID: 10959988.
    Abstract:
    In miniaturized separation techniques, such as capillary electrophoresis (CE) or capillary liquid chromatography (LC), conventional on-line radioactivity detection of labeled compounds is restricted, because of insufficient sensitivity. It will be shown that a microplate scintillation counter for 96-well plates (TopCount) can be used as a sensitive and easy-to-handle radioactivity detector for capillary LC and CE. The attractive combination of capillary LC, eluent fractionation, and subsequent off-line counting is described. The new method is applied for rapid and sensitive separation and detection of 3H-labeled parent drug and its metabolites at levels between 25 and 700 cpm in rat urine. The advantages of capillary LC coupled to the TopCount, and combined with LC-MS data, can be of benefit in many analytical areas, including the characterization of metabolites at low concentration within complex biological fluids. With the same setup, the fractionation with subsequent off-line counting is equally applicable to CE. This is demonstrated with electrophoretically separated 14C-labeled impurities, nicely resolved from a negatively charged main compound, at low levels.