Search MEDLINE/PubMed
Title: Rapid, simple bioassay for 5-fluorocytosine in the presence of amphotericin B. Author: Kaspar RL, Drutz DJ. Journal: Antimicrob Agents Chemother; 1975 Apr; 7(4):462-5. PubMed ID: 1096801. Abstract: It is important that serum levels of 5-fluorocytosine (5FC) be measured to insure therapeutic levels while avoiding toxicity. This is particularly true in patients with renal insufficiency. The concurrent use of amphotericin B and 5FC complicates the measurement of 5FC in the serum, since fungi used in conventional bioassay systems are uniformly susceptible to amphotericin B. This paper describes the development of a simple, reliable, 6-h bioassay for 5FC in the presence of amphotericin B. The assay is based upon the fact that 5FC diffuses readily through yeast nitrogen base agar, whereas amphotericin B apparently does not. This assay allows rapid adjustments in therapy of patients receiving both 5FC and amphotericin B and has permitted us to maintain 5FC serum levels between limits of 25 and 120 mug/ml in patients with altered renal function.[Abstract] [Full Text] [Related] [New Search]