  • Title: A novel balloon angioplasty catheter impregnated with beta-particle emitting radioisotopes for vascular brachytherapy to prevent restenosis; first in vivo results.
    Author: Hehrlein C, Kovacs A, Wolf GK, Yue N, Nath R.
    Journal: Eur Heart J; 2000 Dec; 21(24):2056-62. PubMed ID: 11102256.
    Abstract:
    BACKGROUND: According to early clinical trials, vascular brachytherapy performed prior to or shortly after angioplasty is very effective in reducing restenosis rates. The purpose of this study was to investigate the effects of a novel radioactive catheter that allows simultaneous balloon angioplasty and beta-particle irradiation in the prevention of restenosis. MATERIAL AND METHODS: The balloon surface of an angioplasty catheter was impregnated with the radioisotope(32)P. Dosimetry calculations using a Monte Carlo method were performed at a radial distance of 0.2 mm from the balloon surface. Rabbit iliac arteries were dilated and simultaneously irradiated with a dose of 20 Gy delivered to the adventitia. Control arteries were only dilated and not irradiated. Neointimal areas, cell numbers and the perimeter of the arteries were measured by histomorphometry after 6 weeks. RESULTS: Neointima formation was reduced after balloon dilatation and simultaneous beta-particle irradiation using the(32)P impregnated angioplasty catheter as compared to balloon dilatation alone with a non-impregnated catheter (0.09+/-0.06 vs 0.27+/-0.09 mm(2)neointimal area and 168+/-45 vs 360+/-133 cells/0.05 mm(2)neointima, P<0.001 vs control, respectively). In addition, balloon dilatation with the(32)P impregnated angioplasty catheter increased the vessel perimeter as compared to balloon dilatation with a non-impregnated catheter (4. 7+/-0.2 vs 3.9+/-0.3 mm, P<0.001 vs control). CONCLUSIONS: Simultaneous balloon dilatation and vascular brachytherapy with a novel(32)P impregnated angioplasty catheter markedly reduces restenosis in vivo by preventing neointimal hyperplasia and constrictive vascular remodelling.
    [Abstract] [Full Text] [Related] [New Search]