These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Continuous measurement of core body temperature in preterm infants.
    Author: Dollberg S, Rimon A, Atherton HD, Hoath SB.
    Journal: Am J Perinatol; 2000; 17(5):257-64. PubMed ID: 11110344.
    Abstract:
    We tested a transcutaneous core temperature sensor using a method that relies on the principle of zero heat flow. We tested the hypothesis that transcutaneous and rectal temperatures would track within 0.3 degrees C of each other for >90% of the time. A thermistor was placed between the infant's abdomen or back and the incubator's or radiant warmer's mattress, or within the axilla, attached to the skin with a foam adhesive disk insulator. Thirty preterm infants were either placed on their abdomens or backs in a convective incubator or under a radiant warmer, and continuous transcutaneous and rectal temperatures were measured for 1 hour. There were no significant differences between abdominal and core temperatures or between axillary and core temperatures measured in double-walled convective incubators or in radiant warmers. The rectal-abdominal temperature difference was significantly less than the rectal-axillary difference (p < 0.02) in convective incubators, but not when the infant was placed prone under radiant warmers (p = 0.27). Transcutaneous thermometry is reliable for monitoring core body temperature as indicated by rectal temperature in stable preterm infants in a convective incubator.
    [Abstract] [Full Text] [Related] [New Search]