These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: An outbreak of Pseudomonas aeruginosa infection associated with contaminated urodynamic equipment.
    Author: Yardy GW, Cox RA.
    Journal: J Hosp Infect; 2001 Jan; 47(1):60-3. PubMed ID: 11161900.
    Abstract:
    Seven patients developed Pseudomonas aeruginosa urinary tract infections following urodynamic studies, over a two-month period. One patient developed septicaemia and meningitis and died following rupture of a berry aneurysm. Two others required hospital admission for intravenous antibiotic treatment. Pseudomonas aeruginosa was isolated from a pressure dome which covered the pressure transducer of the urodynamic system used for assessing bladder pressure. The device packaging carried the symbol designating this as a 'single use' product, but as an economy measure, a local decision had been taken to change the device monthly.