  • Title: Direct determination of lead in biological samples by electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICP-MS) after furnace-fusion in the sample cuvette-tungsten boat furnace.
    Author: Okamoto Y.
    Journal: Fresenius J Anal Chem; 2000 Jun; 367(3):300-5. PubMed ID: 11227463.
    Abstract:
    The newly conceived electrothermal vaporization (ETV) system using a tungsten boat furnace (TBF) sample cuvette was designed for the direct analysis of solid samples with detection by inductively coupled plasma mass spectrometry (ICP-MS). Into this small sample cuvette, a solid mixture of the biological samples and diammonium hydrogenphosphate powder as a fusion flux was placed and situated on a TBF. Tetramethylammonium hydroxide solution was added to the mixture. After the on-furnace digestion had been completed, the analyte in the cuvette was vaporized and introduced into the ICP mass spectrometer. The solid samples were analyzed by using a calibration curve prepared from the aqueous standard solutions. The detection limit was estimated to be 5.1 pg of lead, which corresponds to 10.2 ng g(-1) of lead in solid samples when a prepared sample amount of 1.0 mg was applied. The relative standard deviation for 8 replicate measurements obtained with 100 pg of lead was calculated to be 6.5%. The analytical results for various biological samples are described.