These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: B1 field homogeneity and SAR calculations for the birdcage coil. Author: Ibrahim TS, Lee R, Baertlein BA, Robitaille PM. Journal: Phys Med Biol; 2001 Feb; 46(2):609-19. PubMed ID: 11229737. Abstract: The finite-difference time-domain (FDTD) method is used to model a birdcage resonator. All the coil components, including the wires, lumped capacitors and the source, are geometrically modelled together. As such, the coupling effects within the birdcage, including the interactions of coil, source and human head, are accurately computed. A study of the transverse magnetic (B1) field homogeneity and the specific absorption rate (SAR) is presented on an anatomically detailed human head model at 64 and 200 MHz representing 1.5 and 4.7 T MRI systems respectively. Unlike that at 64 MHz, the B1 field distribution is found to be inhomogeneous at 200 MHz. Also, high local SAR values are observed in the tissue near the source due to the coupling between the source and the head at 200 MHz.[Abstract] [Full Text] [Related] [New Search]