These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [A validated finite element model of the human spine--description of the model and initial application]. Author: Pitzen T, Matthis D, Steudel WI. Journal: Z Orthop Ihre Grenzgeb; 2001; 139(1):40-4. PubMed ID: 11253521. Abstract: AIM: To generate a finite-element model of the human cervical spine and evaluate the first application of the model to the analysis of new c-spine implants. METHODS: CT-data were used to generate a three-dimensional, anisotrophic, linear model of the human C4-C7 motion segments using the software ANSYS 5.4. As a next step, anterior cervical fusion and plate fixation using mono- and bicortical screws was simulated in the model. Loading of the finite-element models was simulated using pure moments of +/- 2.5 Nm in flexion/extension, axial left/right rotation, and left/right lateral bending. The range of motion was calculated. The results were compared to the results of an in vitro study using human cadaveric c-spine segments C4-C7, with the same implants and moments on both the intact and surgically treated specimens. RESULTS: The results obtained by the finite-element model were always within one standard deviation of the results of the in vitro study. CONCLUSION: Keeping in mind the simplifications of such a mathematical model, it may be used for a first analysis of the shape of new c-spine implants or to predict the initial stability of a new device.[Abstract] [Full Text] [Related] [New Search]