These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Clinical evaluation of the Shin-Nippon SRW-5000 autorefractor in children. Author: Chat SW, Edwards MH. Journal: Ophthalmic Physiol Opt; 2001 Mar; 21(2):87-100. PubMed ID: 11261351. Abstract: The Canon Autoref R-1 is an 'open-field' autorefractor which has been widely used for research purposes for the past 20 years, but is no longer manufactured. A new autorefractor, the Shin-Nippon SRW-5000, is now available, and if measures using this instrument are shown to be equally accurate and reliable, is likely to replace the R-1. Here we report on the accuracy and reliability (repeatability and reproducibility) of refraction measures in a paediatric population (from 4 to 8 years of age). Subject numbers were 44 for cycloplegic measures and 53 for non-cycloplegic measures. As would be expected, agreement with cycloplegic refraction and reliability were better when SRW-5000 measures were taken using cycloplegia. Repeatability results from the SRW-5000 autorefractor, both with and without cycloplegia were similar to those reported for the Canon R-1.[Abstract] [Full Text] [Related] [New Search]