These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Rapid and sensitive biosensor for Salmonella. Author: Pathirana ST, Barbaree J, Chin BA, Hartell MG, Neely WC, Vodyanoy V. Journal: Biosens Bioelectron; 2000 Jun; 15(3-4):135-41. PubMed ID: 11286330. Abstract: The rapid and sensitive detection of Salmonella typhymurium based on the use of a polyclonal antibody immobilized by the Langmuir-Blodgett method on the surface of a quartz crystal acoustic wave device was demonstrated. The binding of bacteria to the surface changed the crystal resonance parameters; these were quantified by the output voltage of the sensor instrumentation. The sensor had a lower detection limit of a few hundred cells/ml, and a response time of < 100 s over the range of 10(2)-10(10) cells/ml. The sensor response was linear between bacterial concentrations of 10(2)-10(7) cells/ml, with a sensitivity of 18 mV/decade. The binding of bacteria was specific with two binding sites needed to bind a single cell. The sensors preserve approximately 75% of their sensitivity over a period of 32 days.[Abstract] [Full Text] [Related] [New Search]