These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Automated urinalysis. Evaluation of the Sysmex UF-50. Author: Okada H, Sakai Y, Kawabata G, Fujisawa M, Arakawa S, Hamaguchi Y, Kamidono S. Journal: Am J Clin Pathol; 2001 Apr; 115(4):605-10. PubMed ID: 11293910. Abstract: We assessed the Sysmex UF-50 for reproducibility of results and carryover rate by performing between- and within-run precision analyses on 315 urine samples, evaluated the feasibility of using the UF-50 to measure urinary cellular and noncellular components by comparing results from the UF-50 with results of manual urinalysis using the Kova system, and performed side-by-side comparison of the within-run reproducibility from the UF-50, the UF-100, and the Kova system. Results from the UF-50 and UF-100 were highly reproducible, and the carryover rate was 0.5% or less for the urinary components. In between-run precision assays, the coefficients of variation for UF-50 results for all cellular components were less than 10%. The agreement (gamma statistics) between values from the UF-50 and the Kova system was excellent for RBC, WBC, and bacterial counts. The cell counts from the UF-50 for RBCs, WBCs, epithelial cells, and bacteria were 52%, 63%, 54%, and 110%, respectively, of those measured by manual urinalysis. The UF-50 performed quantitative analysis in 72 seconds, compared with 330 seconds for manual methods. The UF-50 is suitable for the first screening to detect hematuria, pyuria, and bacteriuria.[Abstract] [Full Text] [Related] [New Search]