  • Title: Comparative studies on the certification of reference materials by ICPMS and TIMS using isotope dilution procedures.
    Author: Klingbeil P, Vogl J, Pritzkow W, Riebe G, Müller J.
    Journal: Anal Chem; 2001 Apr 15; 73(8):1881-8. PubMed ID: 11338606.
    Abstract:
    A comparison of different isotope dilution mass spectrometric (IDMS) procedures using inductively coupled plasma mass spectrometry (ICPMS) and thermal ionization mass spectrometry (TIMS) was carried out to examine the degree of equivalence between the used procedures in terms of requirements for reference material certification. The comparison was based on the measurement results and their uncertainties. The sample used in this study is a pure zinc metal to be certified by the Bureau Communie de Référence (BCR) for amount contents of different trace elements. This study focuses on cadmium and thallium. The TIMS values contributed to the certified values. To guarantee identical conditions as far as possible for the procedures under investigation, the samples were split into subsamples after spiking and digestion took place. Thus, every IDMS procedure started with an identical set of samples. In total, four different IDMS procedures and one external calibration procedure using internal standardization as an example of routine analysis were applied. The IDMS procedures divide in a group with and a group without trace/matrix separation. Multicollector TIMS (TI-MC-MS) and multicollector ICPMS (ICP-MC-MS) were used in combination with trace/matrix separation, whereas quadrupole ICPMS (ICP-QMS) and ICP-MC-MS were also applied to nonseparated samples. All IDMS results agree well within their combined uncertainties, while some results from the external calibration procedure do not. IDMS results obtained by ICPMS without separation are comparable to those obtained by TI-MC-MS with separation regarding precision and accuracy. The smallest uncertainties were achieved using ICP-MC-MS in combination with trace/matrix separation.
    [Abstract] [Full Text] [Related] [New Search]