Search MEDLINE/PubMed
Title: Prothrombin time ratio is reduced by magnesium contamination in evacuated blood collection tubes. Author: van den Besselaar AM, van Dam W, Sturk A, Bertina RM. Journal: Thromb Haemost; 2001 Apr; 85(4):647-50. PubMed ID: 11341499. Abstract: Magnesium ions were detected in sodium citrate solutions in several lots of evacuated blood collection tubes. The mean concentrations ranged between 1.3 and 1.6 mmol/L. Magnesium was also present in the rubber stoppers of the blood collection tubes and could be leached into the citrate solution. It was shown that magnesium added to citrated plasma shortened the prothrombin time of both coumarin and normal plasma. The effect of magnesium was relatively greater on coumarin than on normal plasma resulting in reduced prothrombin time ratio. Shortening of the prothrombin time was also observed when magnesium chloride was added to dialysed plasma, i.e., in the absence of citrate. These results indicate that magnesium contamination can interfere with accurate INR determination in the control of oral anticoagulant therapy.[Abstract] [Full Text] [Related] [New Search]