These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Simultaneous determination of norepinephrine and epinephrime in plasma by using high performance liquid chromatograph]. Author: Zou Y, Qin Y, Liang M, Yu Q, Huang Y. Journal: Hua Xi Yi Ke Da Xue Xue Bao; 1999 Dec; 30(4):457-9. PubMed ID: 11387970. Abstract: A sensitive and rapid method was reported for the determination of nerepinephrine (NE) and epinephrine (E) in plasma by using high-performance liquid chromatography with an electrochemical detector (HPLC-EC) and YWG-C18 column(5u). 3,4-Dihydroxybenzylamine(DHBA) was used as an internal standard. The mobile phase of 0.05 mol/L actate buffer(pH3.0)-methanol-0.05 mol/L EDTA-Na2(78:20:2) with of 200 mg/L SDS solution was pumped through the column with a flow rate of 1.0 ml/min. The lengths of retention time for NE, E and DHBA(IS) was 7.3 min, 8.3 min and 10.4 min, respectively. The standard curve was linear within the concentration range of 31.25 to 4000 ng/L for NE and E (NE: r = 0.9999 E: r = 0.9996). The within day RSD (%) were less than 1.87% for NE and less than 2.07% for E, inter-day RSD(%) were less than 8.92% for NE and less than 9.65% for E respectively. The average extraction recovery(%) for NE and E were 93.5% +/- 5.50% and 91.3% +/- 1.77% respectively. This method has been applied to the diagnosis of pheochromocytoma and clinical research.[Abstract] [Full Text] [Related] [New Search]