These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: High-performance liquid chromatographic assay of zonisamide in human plasma using a non-porous silica column. Author: Nakamura M, Hirade K, Sugiyama T, Katagiri Y. Journal: J Chromatogr B Biomed Sci Appl; 2001 May 05; 755(1-2):337-41. PubMed ID: 11393722. Abstract: A new method for measuring zonisamide (ZNS) in plasma by high-performance liquid chromatography was developed by using a 2-microm reversed-phase non-porous silica column. ZNS in plasma was first purified with a column extraction technique and injected onto the non-porous silica column. Calibration curve was linear over the concentration range of 1-80 microg/ml in plasma. The recoveries of ZNS added to plasma were more than 95.4% with the coefficient of variation less than 9.0%. We developed a rapid routine method using the non-porous silica column that was accurate and improved solvent consumption in the measurement of ZNS.[Abstract] [Full Text] [Related] [New Search]