These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Complications in pacemaker treatment caused by electrodes (author's transl)]. Author: Nordeck E, Buckesfeld R, Kirchhoff PG, Neubaur J. Journal: Dtsch Med Wochenschr; 1975 Jun 06; 100(23):1282, 1285-8. PubMed ID: 1140085. Abstract: A total of 531 transvenous and 240 thoraco-abdominal electrodes were implanted into 624 patients, the latter as part of a thoracotomy. The complication rate was 35% for transvenous and 33% for epicardial implantations. The most common complication with transvenous electrodes was dislocation (17%), followed by myocardial (5%) and skin perforation (2%). With epicardial electrodes the complications were exit blocks (9%), disconnections (5.5%) and wire breaks (14%). These complication occurred in 1.5%, 2% and 4.5%, respectively, with transvenous electrodes. Electrode dislocations were particularly common in the first year after implantation, while wire breaks became more common as the interval sinceimplantation increased. Overall there was, therefore, no significant difference in complication rate between the two types of electrodes. But the surgical and early mortality ratewas higher after epicardial implanation (with thoracotomy)than transvenous implantation (11% and 6%, respectively). Transvenous implantation is recommended not only for old but also foryoung patients (with a long life expectancy), in combination with long-life pacemakers.[Abstract] [Full Text] [Related] [New Search]