  • Title: Determination of serotonin in microdialysis samples from rat brain by microbore column liquid chromatography with post-column derivatization and fluorescence detection.
    Author: Yoshitake T, Iizuka R, Kehr J, Nohta H, Ishida J, Yamaguchi M.
    Journal: J Neurosci Methods; 2001 Aug 30; 109(2):91-6. PubMed ID: 11513943.
    Abstract:
    The present paper describes a new method for on-line determination of 5-HT in brain microdialysates from awake rats by microbore column liquid chromatography with post-column derivatization and fluorescence detection. The derivatization reagent contained 1 mM benzylamine and 0.5 mM potassium hexacyanoferrate (III), both dissolved in a mixture of acetonitrile and 25 mM borate buffer (pH 11.0) (1:1, v/v). The limit of detection (S/N=3) for 5-HT was 0.5 fmol/20 microl. The samples were injected every 20 min onto a microbore column packed with C18 silica gel. The method exhibits an excellent stability over the periods of at least 12-24 h. The basal levels of 5-HT from 25 awake rats were 7.10+/-1.06 fmol/20 microl in the dorsal hippocampus and 4.64+/-0.91 fmol/20 microl (mean+/-SD) in the striatum. The 5-HT release increased to about 1500% during the perfusion with 100 mM K(+) containing Ringer solution or it was reduced to 60 or 40% during the perfusion with 1 microM tetrodotoxin or calcium free Ringer, respectively. The new method can be used to monitor extracellular 5-HT following acute systemic drug administration.