  • Title: Model-based administration of inhalation anaesthesia. 2. Exploring the system model.
    Author: Lerou JG, Booij LH.
    Journal: Br J Anaesth; 2001 Jan; 86(1):29-37. PubMed ID: 11575406.
    Abstract:
    We explored our model by displaying its new capabilities, testing its sensitivity to variations in input data and illustrating its use. Its multiple-gas character allows simulation of the mechanisms governing concentration and second gas effects. Simulating the volume of a standing bellows makes it possible to test algorithms for automated closed-circuit anaesthesia. Using desflurane, the model's sensitivity to changes in blood/gas partition coefficient (range 0.42-0.576), cardiac output and minute ventilation was analysed. The model was very sensitive to changes in blood solubility; other results agreed with those reported previously. An alveolar isoflurane tension of 1% atm was rapidly attained and maintained, even using 0.5 litres min(-1) of fresh gas, when isoflurane was 'co-administered' through a vaporizer set to 3.5 vol% and a single aliquot (1.25 ml liquid) injected into the expiratory limb. As a result of its credibility and capabilities, the model is to be tested in the clinical setting.