Search MEDLINE/PubMed
Title: Selectivity assessment of DB-200 and DB-VRX open-tubular capillary columns. Author: Kiridena W, Koziola WW, Poole CF. Journal: J Chromatogr A; 2001 Oct 12; 932(1-2):171-7. PubMed ID: 11695864. Abstract: The solvation parameter model is used to study the influence of composition and temperature on the selectivity of two poly(siloxane) stationary phases used for open-tubular capillary column gas chromatography. The poly(methyltrifluoropropyldimethylsiloxane) stationary phase, DB-200, has low cohesion, intermediate dipolarity/polarizability, low hydrogen-bond basicity, no hydrogen-bond acidity, and repulsive electron lone pair interactions. The DB-VRX stationary phase has low cohesion, low dipolarity/polarizability, low hydrogen-bond basicity and no hydrogen-bond acidity and no capacity for electron lone pair interactions. The selectivity of the two stationary phases is complementary to those in a database of 11 stationary phase chemistries determined under the same experimental conditions.[Abstract] [Full Text] [Related] [New Search]