These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Automated chromatographic system for the simultaneous measurement of plasma pregnenolone and 17-hydroxypregnenolone by radioimmunoassay.
    Author: Riepe FG, Wonka S, Partsch CJ, Sippell WG.
    Journal: J Chromatogr B Biomed Sci Appl; 2001 Nov 05; 763(1-2):99-106. PubMed ID: 11710589.
    Abstract:
    A new, simple, rapid and highly practicable automated chromatographic system for the separation, and a sensitive radioimmunoassay system for the subsequent measurement of pregnenolone and 17-hydroxypregnenolone has been developed. Pregnenolone and 17-hydroxypregnenolone were extracted with methylene chloride and separated from cross-reacting steroids by mechanised Sephadex-LH20 multi-column chromatography. Anti-pregnenolone and anti-17-hydroxypregnenolone were obtained by immunising rabbits with pregnenolone-20-oxime-BSA and 17-hydroxypregnenolone-20-oxime-BSA. The lower detection limit of the assay is 0.15 and 0.28 nmol/l for pregnenolone and 17-hydroxypregnenolone, respectively. Normal values for this assay in young male adults, in adult females, and in prepubertal boys and girls were established as a basis for the functional diagnosis of androgen excess syndromes/steroidogenesis defects.
    [Abstract] [Full Text] [Related] [New Search]