These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: High-performance liquid chromatographic procedure for routine residue monitoring of seven sulfonamides in milk. Author: Furusawa N, Kishida K. Journal: Fresenius J Anal Chem; 2001 Dec; 371(7):1031-3. PubMed ID: 11769794. Abstract: A simplified method for routine monitoring of 7 residual sulfonamides (SAs) (sulfadiazine (SDZ), sulfamerazine (SMR), sulfadimidine (SDD), sulfamonomethoxine (SMM), sulfamethoxazole (SMX), sulfadimethoxine (SDM), and sulfaquinoxaline (SQ)) in milk using high-performance liquid chromatography (HPLC) with a photodiode array detector is described. The spiked and blank samples were cleaned up by using an Ultrafree-MC/PL centrifugal ultrafiltration unit. For determination/identification, a Mightysil RP-4 GP column and a mobile phase of 25% (v/v) ethanol in water with a photodiode array detector were used. Average recoveries from milk samples spiked with 0.05, 0.1, 0.2, and 0.5 microg mL(-1) of each drug were >82%. The inter- and intra-assay variabilities were 2.0-3.1%. The practical detection limits for 7 SAs were 0.005-0.02 microg mL(-1). The total time and amount of solvent required for the analysis of one sample were <40 min and <6 mL of ethanol, respectively. No toxic solvents were used.[Abstract] [Full Text] [Related] [New Search]