These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Polyaniline-modified cholinesterase sensor for pesticide determination.
    Author: Ivanov AN, Lukachova LV, Evtugyn GA, Karyakina EE, Kiseleva SG, Budnikov HC, Orlov AV, Karpacheva GP, Karyakin AA.
    Journal: Bioelectrochemistry; 2002 Jan; 55(1-2):75-7. PubMed ID: 11786345.
    Abstract:
    Cholinesterase sensors based on glassy carbon and planar epoxy graphite electrodes modified with processed polyaniline have been developed and examined for pesticide detection. The modification of electrode surface with polyaniline provides high operational stability and sensitivity towards the pesticides investigated. The detection limits found (coumaphos, 0.002, trichlorfon, 0.04, aldicarb, 0.03, methiocarb, 0.08 mg l(-1)) make it possible to detect the pollutants in the waters on the level of limited threshold levels without sample preconcentration.
    [Abstract] [Full Text] [Related] [New Search]