These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Improved transcutaneous bilirubinometry: comparison of SpectR(X) BiliCheck and Minolta Jaundice Meter JM-102 for estimating total serum bilirubin in a normal newborn population.
    Author: Robertson A, Kazmierczak S, Vos P.
    Journal: J Perinatol; 2002 Jan; 22(1):12-4. PubMed ID: 11840236.
    Abstract:
    OBJECTIVE: To compare a new transcutaneous bilirubinometer, which uses multiple wavelength analysis of reflectance data (BiliCheck system), and the commonly used two-wavelength bilirubinometer (Jaundice Meter JM-102) to estimate serum bilirubin. STUDY DESIGN: Jaundiced newborn term infants (101 babies) had transcutaneous bilirubinometry (TCB) using each bilirubinometer, a determination of skin color using a skin color chart, and a total serum bilirubin determination. RESULTS: The variability of the BiliCheck system was less than the variability of the Jaundice Meter JM-102. The accuracy of the BiliCheck system was not affected by skin color, whereas the Jaundice Meter JM-102 was affected. CONCLUSION: Multiple wavelength analysis, as used in the BiliCheck system, is a significant improvement in TCB.
    [Abstract] [Full Text] [Related] [New Search]