These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [An ambulatory respirometer (Hotmate) as a tool for screening of sleep apnea syndrome]. Author: Tanaka M, Katagai K, Tsutsumi Y, Namekawa T, Fukuhara T, Takizawa H, Nakahara K, Ohga E. Journal: Rinsho Byori; 2002 Jan; 50(1):68-73. PubMed ID: 11871139. Abstract: The gold standard diagnostic method for sleep apnea syndrome(SAS) is overnight polysomnography(PSG), but is costly in terms of time and money. We studied the usefulness of a 24-hour ambulatory respirometer equipped with oximeter(Hotmate) for screening of SAS. Seventy-six cases of suspected SAS were enrolled(68 males and 8 females, mean age 51). The correlation between data from Hotmate and PSG was evaluated in 24 cases who underwent both of the tests for the final diagnosis of SAS. There was a good correlation between the two parameters of the data obtained by Hotmate(H) (H-apnea index(AI) vs H-desaturation index(DI)). Among 24 cases who underwent both Hotmate and PSG, there was a good correlation between the data from PSG and Hotmate(PSG-AI vs H-AI: r = 0.80, p < 0.001). Both sensitivity and specificity were highest when screening criteria of H-DI > 15 was utilized(sensitivity = 91.7%, specificity = 66.7%). Our findings suggest that the respiromonitor with oximeter is useful for the screening the patients with SAS.[Abstract] [Full Text] [Related] [New Search]