These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Experimental investigations of image quality in X-ray mammography with conventional screen film system (SFS), digital phosphor storage plate in/without magnification technique (CR) and digital CCD-technique (CCD). Author: Schulz-Wendtland R, Aichinger U, Säbel M, Böhner C, Dobritz M, Wenkel E, Bautz W. Journal: Rontgenpraxis; 2001; 54(4):123-6. PubMed ID: 11883115. Abstract: Comparison of image quality in X-ray mammography between conventional film screen film system (SFS), digital phosphor storage plate in and without magnification technique (CR) and digital CCD-technique (CCD). Radiograms of an RMI-mammography phantom were acquired using a conventional screen film system, three digital storage plate systems and two digital systems in CCD-technique. Additionally, the radiograms of one digital phosphor storage plate system were post-processed regarding contrast and included in the comparison. The detectability of details was best with the digital mammography in CCD-technique. After confirming these promising results in clinical studies, digital mammography should be able to replace conventional screen film technique.[Abstract] [Full Text] [Related] [New Search]