These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Bimanual phacoemulsification through 2 stab incisions. A wound-temperature study.
    Author: Soscia W, Howard JG, Olson RJ.
    Journal: J Cataract Refract Surg; 2002 Jun; 28(6):1039-43. PubMed ID: 12036651.
    Abstract:
    PURPOSE: To determine the thermal parameters of bimanual ultrasound-based phacoemulsification through 2 stab incisions with a bare aspiration needle. SETTING: In vitro laboratory. METHODS: The Sovereign phacoemulsification unit (Allergan) was used in 3 modes in conjunction with the Olson irrigating chopper (ASICO) in 6 fresh human cadaver eye-bank eyes to determine the temperature at which a wound burn would occur. A wound burn was defined as temperature exceeding 45 degrees C, whitening around the wound, or wound contraction. RESULTS: One hundred percent ultrasound power was required to create a wound burn in the unoccluded mode in continuous ultrasound and the burst mode. The time to create the wound burn was 160 and 180 seconds in 2 eyes in continuous ultrasound and 42 and 50 seconds in 2 eyes in the burst mode; it was 58 and 70 seconds in continuous ultrasound with complete occlusion and required 80% power. CONCLUSIONS: The energy necessary to obtain a wound burn during 2-incision bimanual ultrasound-based phacoemulsification with a bare aspiration needle in human cadaver eyes was extremely high. The parameters necessary to create a wound burn in each modality occurred well beyond normal clinical settings and provide evidence that 2-stab ultrasound-based lens extraction can be performed safely with the Sovereign machine.
    [Abstract] [Full Text] [Related] [New Search]