These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Taylor dispersion monitored by electrospray mass spectrometry: a novel approach for studying diffusion in solution. Author: Clark SM, Leaist DG, Konermann L. Journal: Rapid Commun Mass Spectrom; 2002; 16(15):1454-62. PubMed ID: 12125022. Abstract: This work describes a novel approach for monitoring analyte diffusion in solution that is based on electrospray ionization mass spectrometry (ESI-MS). A mass spectrometer at the end of a laminar flow tube is used to measure the Taylor dispersion of an initially sharp boundary between two solutions of different analyte concentration. This boundary is dispersed by the laminar flow profile in the tube. However, this effect is diminished by analyte diffusion that continuously changes the radial position, and hence the flow velocity of individual analyte molecules. The steepness of the resulting dispersion profile therefore increases with increasing diffusion coefficient of the analyte. A theoretical framework is developed to adapt the equations governing the dispersion process to the case of mass spectrometric detection. This novel technique is applied to determine the diffusion coefficients of choline and cytochrome c. The measured diffusion coefficients, (11.9 +/- 1.0) x 10(-10) m(2) s(-1) and (1.35 +/- 0.08) x 10(-10) m(2) s(-1), respectively, are in agreement with the results of control experiments where the Taylor dispersion of these two analytes was monitored optically. Due to the inherent selectivity and sensitivity of ESI-MS, it appears that the approach described in this work could become a valuable alternative to existing methods for studying diffusion processes, especially for experiments on multicomponent systems.[Abstract] [Full Text] [Related] [New Search]