  • Title: Preparation of porous titania sol-gel matrix for immobilization of horseradish peroxidase by a vapor deposition method.
    Author: Yu J, Ju H.
    Journal: Anal Chem; 2002 Jul 15; 74(14):3579-83. PubMed ID: 12139071.
    Abstract:
    A new and facile vapor deposition method has been developed for the preparation of sol-gel matrix. This method was used to form a titania sol-gel thin film and to immobilize horseradish peroxidase (HRP) on a glassy carbon electrode surface for the production of an amperometric hydrogen peroxide biosensor. This process prevented the cracking of conventional sol-gel-derived glasses. The morphologies of both titania sol-gel and the enzyme membranes were characterized using scanning electron microscopy and proved to be chemically clean, porous, and homogeneous and to have a very narrow particle size distribution. The sol-gel-derived titania-modified electrode retained the enzyme bioactivity and provided for long-term stability of the enzyme in storage. In the presence of catechol as a mediator, the sensor exhibited a rapid electrocatalytic response (less than 5 s), a linear calibration range from 0.08 to 0.56 mM with a detection limit of 1.5 microM and a high sensitivity (61.5 microA mM(-1)) for monitoring of H2O2. Effects of pH and operating potential were also explored for optimum analytical performance by using the amperometric method. The apparent Michaelis-Menten constant of the encapsulated HRP was 1.89 +/- 0.21 mM.