These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Characterisation of an ultra-miniature counter for microdosimetric measurements in a therapeutic 400 MeV/A carbon beam. Author: Endo S, Takada M, Ishikawa M, Hoshi M, Uehara S, Yamaguchi H, Kanai T, Matsufuji N, Shizuma K, Onizuka Y. Journal: Radiat Prot Dosimetry; 2002; 99(1-4):421-4. PubMed ID: 12194345. Abstract: Single event spectra of a clinical carbon beam have been measured by an ultra-miniature tissue-equivalent proportional counter (UMC). In order to cover the energy range of the Bragg peak, the incident energy of the carbon beam was degraded by aluminium plates. Single event spectra for carbon-events incident to the UMC were analysed and selected at several carbon energies using thin scintillation counters. It was found that the dose weighted lineal energy distributions have a doublet peak structure due to incident carbon beam and fragment contributions.[Abstract] [Full Text] [Related] [New Search]