These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: An accurate amperometric glucose sensor based glucometer with eliminated cross-sensitivity. Author: Bani Amer MM. Journal: J Med Eng Technol; 2002; 26(5):208-13. PubMed ID: 12487712. Abstract: This paper represents an experimental analysis of cross-sensitivity, which is the main problem that usually faces amperometric glucose sensors. The elimination of cross-sensitivity is based on experimental determination and verification of new correction factors. The correction factors obtained were then used to design a new glucometer with cross-sensitivity of the amperomtric glucose sensor to urea concentration in human blood eliminated. The accuracy of this glucometer satisfies the clinical requirements and the results of glucose measurements obtained using this glucometer are highly correlated with the glucose measurements for the same blood samples obtained using the well-known Yellow Springs Instrument glucose analyser.[Abstract] [Full Text] [Related] [New Search]