Title: Comparison of results for 13 clinical laboratory determinations with three automated analytical systems. Author: Sampson EJ, Derck DD, Demers LM. Journal: Clin Chem; 1976 Mar; 22(3):346-9. PubMed ID: 1253409. Abstract: We evaluated the Abbott Bichromatic Analyzer-100 (ABA-100) for use in the routine clinical chemistry laboratory by examining 13 different determinations that can be performed on the instrument. Results with the Du Pont "aca" and Technicon continuous-flow systems were compared to the ABA-100 in terms of upper limits of linearity, inter-run coefficient of variation, and results for samples from patients. The upper limits of linearity for the methods on the ABA-100 exceeded all of those for the continuous-flow systems, except for urea nitrogen. Precision of the ABA-100 was as good as or better than that of the aca for all determinations, except for glucose in a normal control serum and creatine kinase and creatinine in an above-normal control serum.[Abstract] [Full Text] [Related] [New Search]