Search MEDLINE/PubMed
Title: Non-invasive cardiac output monitoring by aortic blood flow measurement with the Dynemo 3000. Author: Boulnois JL, Pechoux T. Journal: J Clin Monit Comput; 2000; 16(2):127-40. PubMed ID: 12578070. Abstract: The operating principles and methods for the continuous determination of aortic blood flow (ABF) with the Dynemo 3000 system are described in detail. The system uses a novel transesophageal ultrasonic Echo-Doppler probe simultaneously to measure aortic diameter and blood flow velocity at the same anatomic level, in real-time. Non-invasive ABF measurement is combined with vital sign data from standard monitors to provide a composite hemodynamic profile including volume, afterload and contractility data used by the physician to optimize therapy. A review of the clinical validation and comparison to thermodilution measurements showing a significant positive correlation over a wide range of clinical flow situations is also briefly presented.[Abstract] [Full Text] [Related] [New Search]