These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Non-invasive cardiac output monitoring by aortic blood flow measurement with the Dynemo 3000.
    Author: Boulnois JL, Pechoux T.
    Journal: J Clin Monit Comput; 2000; 16(2):127-40. PubMed ID: 12578070.
    Abstract:
    The operating principles and methods for the continuous determination of aortic blood flow (ABF) with the Dynemo 3000 system are described in detail. The system uses a novel transesophageal ultrasonic Echo-Doppler probe simultaneously to measure aortic diameter and blood flow velocity at the same anatomic level, in real-time. Non-invasive ABF measurement is combined with vital sign data from standard monitors to provide a composite hemodynamic profile including volume, afterload and contractility data used by the physician to optimize therapy. A review of the clinical validation and comparison to thermodilution measurements showing a significant positive correlation over a wide range of clinical flow situations is also briefly presented.
    [Abstract] [Full Text] [Related] [New Search]