These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of Cobas Integra 800 under simulated routine conditions in six laboratories.
    Author: Redondo FL, Bermudez P, Cocco C, Colella F, Graziani MS, Fiehn W, Hierla T, Lemoël G, Belliard A, Manene D, Meziani M, Liebel M, McQueen MJ, Stockmann W.
    Journal: Clin Chem Lab Med; 2003 Mar; 41(3):365-81. PubMed ID: 12705348.
    Abstract:
    The new selective access analyser Cobas Integra 800 from Roche Diagnostics was evaluated in an international multicentre study at six sites. Routine simulation experiments showed good performance and full functionality of the instrument and provocation of anomalous situations generated no problems. The new features on Cobas Integra 800, namely clot detection and dispensing control, worked according to specifications. The imprecision of Cobas Integra 800 fulfilled the proposed quality specifications regarding imprecision of analytical systems for clinical chemistry with few exceptions. Claims for linearity, drift, and carry-over were all within the defined specifications, except urea linearity. Interference exists in some cases, as could be expected due to the chemistries applied. Accuracy met the proposed quality specifications, except in some special cases. Method comparisons with Cobas Integra 700 showed good agreement; comparisons with other analysis systems yielded in several cases explicable deviations. Practicability of Cobas Integra 800 met or exceeded the requirements for more than 95% of all attributes rated. The strong points of the new analysis system were reagent handling, long stability of calibration curves, high number of tests on board, compatibility of the sample carrier to other Roche systems, and the sample integrity check for more reliable analytical results. The improvement of the workflow offered by the 5-position rack and STAT handling like on Cobas Integra 800 makes the instrument attractive for further consolidation in the medium-sized laboratory, for dedicated use of special analytes, and/or as back-up in the large routine laboratory.
    [Abstract] [Full Text] [Related] [New Search]